The latest news and product announcements. Power/surge protection, conditioning, control, battery backup and voltage regulation products. White Papers-Articles On Understanding and Achieving High Availability & High Quality Power Double Conversion True On-Line Battery Backup - Uninterruptible Power Systems & Single Conversion UPS Outdoor UPS for Extreme Temperature/Weather Environmental Conditions In Exterior Rated NEMA Enclosures AC Automatic Transfer Switches For Redundant Power Sources or UPS Bypass AC Power Conditioners - Low-Impedance Isolation Transformer and Ferroresonant Transformers Left navigation bar for the CareBase website focusing on mission-critical power systems and surge protection, utilizing UPS, precision automatic voltage regulators, line conditioners, battery backups, automatic transfer switches, on-line outdoor UPS, surge protective devices and more.
 

CareBase
TM
 

 

 

FOR IMMEDIATE RELEASE

Medical isolation transformer, UL 60601-1, EI core, less than 50 microamps leakage current, neutral-ground bonded output, low-impedance, for patient vicinity equipment.

Contact:
Michael Gibson

CareBase
970.945.2770
michael.gibson@carebase.com 


                                                                                   ILc-1400Med4 shown

ILc-MED SERIES MEDICAL ISOLATION TRANSFORMERS 
UL 60601-1 listed
"low leakage current" ILc-Med Series units enhance medical systems operation and minimize re-booting

Antigo, WI, USA ~ TSi Power Corporation announces its ILc-Med Series of UL 60601-1 listed medical isolation transformers. This series satisfies the requirement for low leakage current in an isolation transformer protecting medical systems located within the vicinity of a patient.

Mr. Peter Nystrom, President of TSi Power Corp., says, "These units not only provide coordinated surge protection of critical medical systems per IEC 61312.3, they also improve the accuracy of medical diagnostic and monitoring equipment. Computerized systems all look to ground for a zero volt logic reference. The ILc-Med Series has a neutral-ground bonded output which eliminates common-mode noise and provides a reliable logic reference."

True isolation plus common-mode noise elimination

The ILc-Med Series contributes to enhanced systems operation, greater reliability, improved repeatability, fewer re-boots or systems lock-ups, and extended life. In addition, a reduction in hospital medical equipment service costs can be expected.

The integral isolation transformer provides complete isolation between the primary and secondary, and permits bonding the output neutral to ground, which completely eliminates all disturbances between neutral and ground, regardless of source. The series inductance of the isolation transformer in combination with a capacitive element and an MOV, provide noise filtering, as well as coordinated multi-stage surge protection.

In addition, the ILc-Med has an extremely low input leakage current rating (typically 50 µA), which exceeds both domestic US (UL 60601-1), Canadian (cUL 60601-1), and international (IEC 60601-1) medical standards for a leakage current limit of 300 µA maximum for non-patient connected medical equipment and instruments.

About TSi Power Corporation

TSi Power manufactures ac line conditioners (including the medical grade series), isolation transformers, ac automatic voltage regulators, automatic transfer switches (ATS), uninterruptible power supplies (UPS), outdoor wide temperature UPS, outdoor ac to dc wide temperature UPS, dc to ac inverters and overvoltage protection devices.

TSi Power specializes in custom products for relatively small volumes. These are created from combining our regular products with our library of various standard board designs and an extensive library of CAD designs. This combination allows us to respond quickly and economically to many varying requirements.

For more information on TSi’s ILc-Med Series, please visit our isolation transformer page.

###